Air bubbles removal technique from intravenous tubing
Downloads
Background: The air bubbles in the intravenous tubing can be fatal if it enters the body. It can result in an air embolism. Air embolism is a condition in which air bubbles enter the blood vessels. Several methods can be used to remove air bubbles in the intravenous tubing. One is replacing it with new intravenous tubing, but this method is considered less efficient. This article describes effective and efficient ways to remove air bubbles from intravenous tubing and tests them for effectiveness.
Technique: Four techniques can be applied when the intravenous tubing is filled with air, namely the roll-up technique, flick technique, combination roll-up and flick technique, and push-pull technique.
Conclusion: From the test, the push-pull technique is the most effective and efficient in removing air from intravenous tubing compared to the others.Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution-NonCommercial 4.0 International License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.